MDS1088 two wire intelligent temperature transmitter

 Summary

MDS1088 is a HART type two wire temperature Transmitter That can be used to detect the thermal resistance, thermocouple, resistance and MV signal, the output of 4-20mA and standard HART signal, and has excellent performance, reliability and anti-interference ability; widely used in various industrial applications.

 One thousand and eighty-eight

Characteristic 

The general thermal resistance, thermocouple
Can be through a variety of Hart equipment configuration
The input and output electromagnetic isolation
The built-in high precision cold end compensation
Dynamic configuration data - storage
24 - bit digital conversion based on digital linearization technique, high precision
The vacuum pouring process, earthquake proof
The electromagnetic compatibility with EN 61326 and EN21
- suitable for installation in the standard junction box (DIN B).
 

The input signal and the basic error


MDS1088

Input range

The basic error

Input type

measuring range

Minimum range

absolute error

Spanerror

heat

electric

Resistance

Pt 100

-200 ~Eight hundred and fiftyC

TwentyC

. Zero point two C

+ 0.1%F.S

Pt 1000

-200 ~Eight hundred and fiftyC

TwentyC

. Zero point three C

+ 0.1%F.S

Cu50

-50 ~One hundred and fiftyC

TwentyC

. Zero point two C

+ 0.1%F.S

Cu100

-50 ~One hundred and fiftyC

TwentyC

. Zero point two C

+ 0.1%F.S

Ni1000

-60 ~Two hundred and fiftyC

TwentyC

. Zero point three C

+ 0.1%F.S

Linear resistance

0-400.

10.

0.1.

+ 0.1%F.S

0-4000.

100.

0.25.

+ 0.1%F.S

heat

electric

even

K type

-200 ~One thousand three hundred and seventy-two C

Fifty C

. Zero point five C

+ 0.1%F.S

S type

Zero ~One thousand seven hundred and sixty-eight C

Two hundredC

. One point five C

+ 0.1%F.S

E type

-200 ~One thousandC

Fifty C

. Zero point five C

+ 0.1%F.S

B type

One hundred~ One thousand eight hundred and twenty C

Two hundredC

. One point five C

+ 0.1%F.S

R type

Zero ~One thousand seven hundred and sixty-eight C

Two hundredC

. One point five C

+ 0.1%F.S

T type

-200 ~Four hundredC

Fifty C

. Zero point five C

+ 0.1%F.S

J type

-200 ~One thousand and two hundredC

Fifty C

. Zero point five C

+ 0.1%F.S

N type

-200 ~One thousand and three hundredC

Fifty C

. One C

+ 0.1%F.S

Millivolt

-10 ~100mV

5mV

30uV

+ 0.1%F.S


Remarks:

1, the precision of test environment temperature 25 - 2 DEG C, 24V power supply, 250 ohm load;
The maximum value of 2, the actual application error and absolute error of range error;

work environment

Working temperature: -40 to 85 DEG C
Storage temperature: -50 to 120 DEG C
Relative humidity: 0 ~ 95%RH allows setting
Atmospheric pressure: 86 ~ 106kPa

The working power supply

Supply voltage: 12 ~ 40V
The power ripple: Vp-p < 1%
Supply voltage: F.S < 0.002% /V effect

insulation

Insulation resistance: 500M ohm (500VAC)
Dielectric strength: 1500VAC/60s (leakage current less than 1.0mA)

input

Input signal: thermal resistance, thermocouple, resistance and MV signal
See the specific input signal and basic error

output


2 wire 4-20mA and standard HART digital signal output
The linear range of output: 3.9 mA = I = 21.0mA
Low fault alarm value: 3.5 ~ 3.8mA (default 3.8mA, user configurable)
High fault alarm value: 21 ~ 22.0mA (default 22mA, user configurable)
Output: Vp-p = 20mV

The basic error

The basic error : Less than or equal to + 0.1% F.S ( See the specific The input signal and the basic error)

temperature drifttemperature drift : Less than or equal to + 0.005% F.S/ C

Maximum load

The maximum load resistance (supply voltage = - 12V / 0.022A).
When the supply voltage is 24V, the maximum load is 545 ohms
 

Thermal resistance parameters

allow sensor The largest single cable line resistance: 10 ohms
sensor Cable resistance: 0.05 /10 ohm.
The thermal resistance of the excitation current: less than 250uA
Thermal resistance connection: two wire or three wire or four wire input, three wire input default
 

The related parameters of thermocouple

Cold end compensation: internal or external compensation, user configurable, default internal compensation
Internal compensation accuracy: less than or equal to 0.5 DEG C (-40 to 85 DEG C)Damping parameters

0 ~ 60s, user configurable, default is 0.0s

 

Start time

Transmitter When the power supply is switched on after 5 minutes can reach the performance specifications range

 

(response time T Ninety )

The damping parameter is 0, the response time is less than 1s